Effects of dental metal implant number and configuration on PET/CT quantification and image quality: A phantom-based study

dc.contributor.authorAkdoğan, Emine Tuna
dc.contributor.authorCoşkun, Sinem
dc.contributor.authorYeyin, Nami
dc.contributor.authorUslu Beşli, Rabia Lebriz
dc.contributor.authorDemir, Mustafa
dc.date.accessioned2026-10-01T08:55:33Z
dc.date.issued2026
dc.departmentİstanbul Kent Üniversitesi, Fakülteler, Diş Hekimliği Fakültesi, Klinik Bilimler Bölümü
dc.description.abstractIntroduction/Objectives: Dental metal implants can generate artifacts in PET/CT imaging, particularly in the head and neck region, affecting quantitative accuracy. This study evaluated variations in standardized uptake value (SUV), contrast (C), and contrast-to-noise ratio (CNR) across predefined dental implant configurations differing in both implant number and spatial arrangement and assessed their effects on simulated tongue, nasopharyngeal, and laryngeal lesions. Methods: A custom-designed head and neck phantom containing extracted human teeth and up to eight titanium dental implants was used. Simulated tongue, nasopharyngeal, and laryngeal tumors were filled with F-18 FDG. PET/CT scans were acquired using seven implant configurations (0–8 implants) and reconstructed as attenuation-corrected (AC), AC with metal artifact reduction (AC+MAR), and non-attenuation-corrected (NoAC) images. SUV, contrast, and CNR were calculated using standardized ROIs. Results: Compared with the implant-free configuration, tongue tumor SUV decreased by 11% and 9.1% on AC and AC+MAR images, respectively, at eight implants. Nasopharyngeal and laryngeal SUV showed only minimal descriptive change with increasing implant number and configurations on AC+MAR images. Contrast and CNR followed similar trends on AC and AC+MAR images, whereas NoAC images consistently showed substantially lower contrast and CNR regardless of implant number. Discussion: Although MAR reduced implant-induced artifacts and partially attenuated SUV underestimation, it did not eliminate the configuration-dependent trend observed across the predefined implant arrangements. Lesions farther from the implants appeared less affected after MAR; however, this should be regarded as a hypothesis-generating observation from a single phantom experiment, requiring validation in patient studies and phantom models with varying implant-to-lesion distances. Conclusion: Dental implants may affect PET/CT quantification, particularly for lesions adjacent to multiple implants; SUV measurements in these regions should be interpreted with caution.
dc.identifier.citationAkdoğan ET., Coşkun S., Yeyin N., Uslu Beşli RL., Demir M. Effects of dental metal implant number and configuration on PET/CT quantification and image quality: A phantom-based study. Current Medical Imaging, 2026, 22, e15734056519233.
dc.identifier.doi10.2174/0115734056519233260918120340
dc.identifier.issn1875-6603
dc.identifier.orcid0000-0001-8608-0862
dc.identifier.orcid0000-0003-3056-7035
dc.identifier.orcid0000-0003-0262-4020
dc.identifier.orcid0000-0003-1868-3957
dc.identifier.orcid0000-0002-9813-1628
dc.identifier.pmid42786833
dc.identifier.scopus2-s2.0-105053739022
dc.identifier.scopusqualityQ3
dc.identifier.urihttps://www.eurekaselect.com/article/158637
dc.identifier.urihttps://hdl.handle.net/20.500.12780/1970
dc.identifier.volume22
dc.identifier.wosWOS:001904827900001
dc.identifier.wosqualityQ4
dc.indekslendigikaynakPubMed
dc.indekslendigikaynakWeb of Science
dc.indekslendigikaynakScopus
dc.language.isoen
dc.publisherBentham Science Publishers
dc.relation.ispartofCurrent Medical Imaging
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı
dc.rightsinfo:eu-repo/semantics/openAccess
dc.subjectPET/CT image quality
dc.subjectDental implant
dc.subjectMetal artifact
dc.subjectAttenuation correction
dc.subjectMetal artifact reduction (MAR)
dc.titleEffects of dental metal implant number and configuration on PET/CT quantification and image quality: A phantom-based study
dc.typeArticle

Dosyalar

Orijinal paket

Listeleniyor 1 - 1 / 1
Yükleniyor...
Küçük Resim
İsim:
akdogan,eminetuna.pdf
Boyut:
7,02 MB
Biçim:
Adobe Portable Document Format
Açıklama:
Tam Metin / Full Text

Lisans paketi

Listeleniyor 1 - 1 / 1
Yükleniyor...
Küçük Resim
İsim:
license.txt
Boyut:
1,17 KB
Biçim:
Item-specific license agreed upon to submission
Açıklama: