Effects of dental metal implant number and configuration on PET/CT quantification and image quality: A phantom-based study
| dc.contributor.author | Akdoğan, Emine Tuna | |
| dc.contributor.author | Coşkun, Sinem | |
| dc.contributor.author | Yeyin, Nami | |
| dc.contributor.author | Uslu Beşli, Rabia Lebriz | |
| dc.contributor.author | Demir, Mustafa | |
| dc.date.accessioned | 2026-10-01T08:55:33Z | |
| dc.date.issued | 2026 | |
| dc.department | İstanbul Kent Üniversitesi, Fakülteler, Diş Hekimliği Fakültesi, Klinik Bilimler Bölümü | |
| dc.description.abstract | Introduction/Objectives: Dental metal implants can generate artifacts in PET/CT imaging, particularly in the head and neck region, affecting quantitative accuracy. This study evaluated variations in standardized uptake value (SUV), contrast (C), and contrast-to-noise ratio (CNR) across predefined dental implant configurations differing in both implant number and spatial arrangement and assessed their effects on simulated tongue, nasopharyngeal, and laryngeal lesions. Methods: A custom-designed head and neck phantom containing extracted human teeth and up to eight titanium dental implants was used. Simulated tongue, nasopharyngeal, and laryngeal tumors were filled with F-18 FDG. PET/CT scans were acquired using seven implant configurations (0–8 implants) and reconstructed as attenuation-corrected (AC), AC with metal artifact reduction (AC+MAR), and non-attenuation-corrected (NoAC) images. SUV, contrast, and CNR were calculated using standardized ROIs. Results: Compared with the implant-free configuration, tongue tumor SUV decreased by 11% and 9.1% on AC and AC+MAR images, respectively, at eight implants. Nasopharyngeal and laryngeal SUV showed only minimal descriptive change with increasing implant number and configurations on AC+MAR images. Contrast and CNR followed similar trends on AC and AC+MAR images, whereas NoAC images consistently showed substantially lower contrast and CNR regardless of implant number. Discussion: Although MAR reduced implant-induced artifacts and partially attenuated SUV underestimation, it did not eliminate the configuration-dependent trend observed across the predefined implant arrangements. Lesions farther from the implants appeared less affected after MAR; however, this should be regarded as a hypothesis-generating observation from a single phantom experiment, requiring validation in patient studies and phantom models with varying implant-to-lesion distances. Conclusion: Dental implants may affect PET/CT quantification, particularly for lesions adjacent to multiple implants; SUV measurements in these regions should be interpreted with caution. | |
| dc.identifier.citation | Akdoğan ET., Coşkun S., Yeyin N., Uslu Beşli RL., Demir M. Effects of dental metal implant number and configuration on PET/CT quantification and image quality: A phantom-based study. Current Medical Imaging, 2026, 22, e15734056519233. | |
| dc.identifier.doi | 10.2174/0115734056519233260918120340 | |
| dc.identifier.issn | 1875-6603 | |
| dc.identifier.orcid | 0000-0001-8608-0862 | |
| dc.identifier.orcid | 0000-0003-3056-7035 | |
| dc.identifier.orcid | 0000-0003-0262-4020 | |
| dc.identifier.orcid | 0000-0003-1868-3957 | |
| dc.identifier.orcid | 0000-0002-9813-1628 | |
| dc.identifier.pmid | 42786833 | |
| dc.identifier.scopus | 2-s2.0-105053739022 | |
| dc.identifier.scopusquality | Q3 | |
| dc.identifier.uri | https://www.eurekaselect.com/article/158637 | |
| dc.identifier.uri | https://hdl.handle.net/20.500.12780/1970 | |
| dc.identifier.volume | 22 | |
| dc.identifier.wos | WOS:001904827900001 | |
| dc.identifier.wosquality | Q4 | |
| dc.indekslendigikaynak | PubMed | |
| dc.indekslendigikaynak | Web of Science | |
| dc.indekslendigikaynak | Scopus | |
| dc.language.iso | en | |
| dc.publisher | Bentham Science Publishers | |
| dc.relation.ispartof | Current Medical Imaging | |
| dc.relation.publicationcategory | Makale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı | |
| dc.rights | info:eu-repo/semantics/openAccess | |
| dc.subject | PET/CT image quality | |
| dc.subject | Dental implant | |
| dc.subject | Metal artifact | |
| dc.subject | Attenuation correction | |
| dc.subject | Metal artifact reduction (MAR) | |
| dc.title | Effects of dental metal implant number and configuration on PET/CT quantification and image quality: A phantom-based study | |
| dc.type | Article |