dc.date.issued2020en_US
dc.identifier.citationYanikoglu, F., Avci, H., Celik, Z. C., & Tagtekin, D. (July 01, 2020). Diagnostic Performance of ICDAS II, FluoreCam and Ultrasound for Flat Surface Caries with Different Depths. Ultrasound in Medicine and Biology, 46, 7, 1755-1760. DOI: https://doi.org/10.1016/j.ultrasmedbio.2020.03.007en_US
dc.identifier.urihttps://doi.org/10.1016/j.ultrasmedbio.2020.03.007
dc.identifier.urihttps://hdl.handle.net/20.500.12780/178
dc.description.abstractABSTRACT The use of ultrasound for the diagnosis of natural caries lesions with different depths on flat surfaces of permanent molar teeth was studied. A total of 200 freshly extracted permanent molar teeth with natural caries lesions were included. After evaluation with International Caries and Detection Assessment System II (ICDAS II), mineral loss and lesions in the teeth were detected with a fluorescence device, FluoreCam and lesion depths were measured using ultrasound. The teeth were sectioned and examined histologically using a microscope. The intraclass correlation coefficient was 0.97 for the ultrasound system, 0.65 for the ICDAS II classification and 0.59 for FluoreCam (p < 0.001 for all). With an increase in ICDAS II scores, mean ultrasound values, FluoreCam–Size and microscope measurements increased numerically, while FluoreCam–Intensity and FluoreCam–Impact measurements inversely decreased (p < 0.001). There was a high level of agreement between histologic diagnosis and ultrasound. Ultrasound and FluoreCam can be used to detect enamel caries on flat surfaces.en_US
dc.language.isoengen_US
dc.publisherELSEVIERen_US
dc.rightsinfo:eu-repo/semantics/closedAccessen_US
dc.subjectCaries detectionen_US
dc.subjectUltrasounden_US
dc.subjectFluorescenceen_US
dc.subjectInternational Cariesen_US
dc.subjectDetection Assessment Systemen_US
dc.titleDiagnostic performance of ICDAS II, fluorecam and ultrasound for flat surface caries with different depthsen_US
dc.typearticleen_US
dc.contributor.departmentİstanbul Kent Üniversitesi, Fakülteler, Diş Hekimliği Fakültesi, Klinik Bilimler Bölümüen_US
dc.contributor.institutionauthorYanıkoğlu, Funda
dc.identifier.volume46en_US
dc.identifier.issue7en_US
dc.identifier.startpage1755en_US
dc.identifier.endpage1760en_US
dc.relation.journalUltrasound in Medicine & Biologyen_US
dc.relation.publicationcategoryKonferans Öğesi - Uluslararası - Kurum Öğretim Elemanıen_US


Bu öğenin dosyaları:

DosyalarBoyutBiçimGöster

Bu öğe ile ilişkili dosya yok.

Bu öğe aşağıdaki koleksiyon(lar)da görünmektedir.

Basit öğe kaydını göster