Towards a phantom-based quantitative assessment of dose reduction and image-quality trade-offs in half- and full-rotation dental CBCT protocols: A thermoluminescent dosimetry approach with clinical insight

Yükleniyor...
Küçük Resim

Tarih

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Springer Nature

Erişim Hakkı

info:eu-repo/semantics/embargoedAccess

Özet

This study aimed to compare full- (360°) and half-rotation (180°) dental cone-beam computed tomography (CBCT) pro tocols in terms of effective dose (ED) and quantitative image quality for the maxilla and mandible regions. An Alderson Radiation Therapy phantom was imaged using a Hyperion X9 Pro (Cefla, Imola) CBCT device with a 10 × 6 cm field of view. Both anatomical regions were acquired using full- and half-rotation protocols. A total of 67 thermoluminescence dosimeters were positioned in the phantom for ED calculations. Quantitative image-quality assessment was performed using axial slices extracted from each volume, and signal-to-noise ratio (SNRs) and contrast-to-noise ratio (CNRs) were calculated. The EDs were 406.33 µSv (full-rotation) and 208.29 µSv (half-rotation) for the maxilla, and 248.94 µSv (full rotation) and 73.63 µSv (half-rotation) for the mandible. These doses corresponded to dose reductions of 48.74% in the maxilla and 70.42% in the mandible. SNRs and CNRs decreased by 36.69% and 34.74% in the maxilla, and by 49.95% and 50.28% in the mandible, respectively. Dose reduction was more pronounced in the mandible than in the maxilla, and this was accompanied by a greater loss in SNR and CNR. It is concluded that protocol selection should be guided by diagnostic requirements and adherence to the “As Low as Diagnostically Acceptable being Indication-oriented and Patient-specific (ALADAIP) principle.

Açıklama

Anahtar Kelimeler

Cone-beam computed tomography, Dosimetry, Image quality, Anthropomorphic phantom, Radiographic imaging

Kaynak

Radiation and Environmental Biophysics

WoS Q Değeri

Scopus Q Değeri

Cilt

Sayı

Künye

Amasya, H., Günay, O., Kesmezacar, F.F. et al. Towards a phantom-based quantitative assessment of dose reduction and image-quality trade-offs in half- and full-rotation dental CBCT protocols: a thermoluminescent dosimetry approach with clinical insight. Radiat Environ Biophys (2026).

Onay

İnceleme

Ekleyen

Referans Veren